2465B(B050351) »  Viewing PA064459     [Image 18 of 23]  :: Jump To  
  First slide Previous slide      Show/Hide EXIF Info Index Page Start/Stop the slide show      Next slide Last slide  
10V level calibrator into CH1 and CH2.

CH1 fast rise response 810pS.

CH2 fast rise response 810pS.

CH1 leveled 6 Division sine wave @ 50KHz. See next picture as the leveled sine wave frequency is increased. (The small sine wave in the center of the CRT is a reflection off the camaera lens)


Next slide
 
PA064459.JPG - 2560 x 1920 - (1.24MB)
10/6/11 6:29 PM

It's about 3dB down at 480Mhz!

Diagnostics all pass.

This scope still has plenty of life left in it.

All four channels input the cal signal through 10X probes.

Album last updated on 10/6/11 7:34 PM
Powered by Jalbum 8.1 and BluPlusPlus skin